Model
|
YJ-180
|
Analyze Unit
|
Throughput
|
Constant 180T/h (non-ISE)
|
Theory
|
Absorption spectrometry, transmission turbidimetry
|
Methodology
|
End-point \Fixed-time\Kinetic\Ion selective electrode
|
Test according set up, random access
|
Reaction plate
|
Contain 81 reaction cuvettes
|
Maximum reaction time
|
15 minutes
|
Minimum reaction volume
|
150 ul
|
Material of cuvettes
|
Special UV-transparent optical glass
|
Cuvettes use
|
Automatically cuvettes washing
|
Optical Unit
|
Light Source
|
Halogen-tungsten lamp
|
Photometer
|
Reversed optics
|
Wavelength
|
340nm-900nm wavelength option
|
Absorption
|
0-4 Abs
|
Reagent Unit
|
Reagent plate
|
60 reagent positions ,support for single and double reagent, reagent completely open
|
Reagent storage temperature
|
2~10 degree
|
Reagent bottle
|
15ml, 20ml, 30ml, 50ml or 70ml ,dead volume is less than 1.0 ml
|
Reagent volume
|
20-500 ul, step by 1.0 ul
|
Reagent/Sample Probe
|
Automatically checked ;Patent technology liquid level detection;Collision protection probe cleaning:Automatic washing for both inside and outside.
|
Sample Unit
|
Sample plate
|
60 sample positions
|
Sample type
|
Human (animal) serum \plasma\urine etc.
|
Samples cup
|
Support one-off micro cups
|
Sample volume
|
2-100 ul, step by 0.1 ul
|
Sample/reagent probe
|
Automatically checked ;Patent technology liquid level detection;Collision protection probe cleaning:Automatic washing for both inside and outside.
|
Temperature control Unit
|
Heating method
|
Direct thermal solid thermostat closed
|
Temperature equilibrium time
|
<3 minutes
|
Reaction temperature
|
37 ±0.1 degree
|
Temperature fluctuation
|
±0.1 degree
|
Operation Unit
|
Software
|
WindowsXP \2000
|
Database
|
SQL Server
|
Data Storage
|
Automatic store & copy; No time and volume limit; Lifetime preservation
|
Print format
|
Can be set up, offer several standard format
|
Calibration mode
|
Linear curve type: Single-point linear, two-point linear and multi-point linear none linear; Curve type: Logistic-Log 4P,Logistic-Log5p,Spline,index automatically use best calibration curve and standard
|
Quality control methods :
|
Real time, within day & between day
|
Quality control diagram
|
L-J \cumulative\twin plot
|
Status display
|
Real time monitoring of reagent residual volume .Running status of reagent tray, samples tray
|
Adjustment
|
Dynamic and real time display of running status; Intelligent carry over settings to adjust test sequence; original reaction data record; Detail profile of alert messages
|
Setting
|
Operation permission setting, Item setting, cross pollution setting, cuvette blank setting
|
Others
|
Water consumption : < 6L/h
|
Power Supply :AC 220V±22V 50Hz±1Hz
|
Frequency :Maximum 600VA
|
Dimension :782mm×581mm×545mm
|
Net Weight : 60kg
|
Accessory :Software
|