Fully Auto Elisa Reader
picture
|
specification
|

|
Measurement System : 8-Channel optical system
Measurement Range : 0.001~4.000Abs
Photometric Accuracy : ±0.01Abs (0.001~1.500Abs)
Repeatability : <±0.5% (0.001~1.500Abs)
Stability : <0.005Abs ( some are 0.007 Abs)
Linearity : <±1% (0.001~1.500Abs)
Filters : Standard:4 filters (405,450,492,630nm) ,other wave length can be optional.
Totally 10 filters can be installed.
Full One year Warranty
|
Function and Technical Specifications
Fundamental parameters
1) Range of working wavelength is from 400nm to 800nm;( some manufacture are 400 ~700 nm,the ranger is less than ours)
2) Range of detection is from 0.000A to 4.000A;
3) Resolving power:0.001 A;
4) Filters when leaving factory:405nm,450nm,492nm,630 nm, with the maximum array of 10 filters between 400nm-800nm simultaneously.
5) Reading speed:single wave :5s,dual waves :10s;
6) Dimension: 420×325×210 mm(l×h×w);
7) Weight: 9 kg;others are 15 kg, the higher weight, the more space and higher freight .
8)Monitor:240*128mm LCD;
9)Printer:external printer.
Ⅳ Technical Specifications
(1) Characters of the filters
Table 1: Characters of the Filters
Accuracy of wavelength (nm)
|
Half-width(nm)
|
±3.0
|
≤12
|
(2 ) The errors of absorbance accuracy are as follows:
Table 2: Absorbance Accuracy
Absorbance (A)
|
Accuracy Error
|
0.000-2.000
|
±0.01
|
>2.000-3.000
|
±0.05
|
>3.000-4.000
|
±0.15
|
(4) The Linearity Errors are as follows:
Table 3: Linearity of the Instrument
Absorbance (A)
|
Error (%)
|
0.000-2.000
|
±1.0%
|
>2.000-3.000
|
±2.0%
|
>3.000-4.000
|
±5.0%
|
(5) Reproducibility Error: ≤0.5%;
(6) Stability Error: ≤±0.005A/10min;